Language
DPS65B1-XXXYYYSPA
Date:2023-01-29    Hits:343
  • Power:65W
  • Output voltage:11.0-23.1V
  • Output current:0.01-5.00A

TUV-PSE.jpg  1674961567788815.gif 1674961567562554.gifCB.gif

Specification Download

Medical & household safety approvals

2MOPP input to output isolation

Touch current ≤100μA

Energy efficiency Level VI

≤0.15W standby power

11V to 24V outputs,up to 65W


Input

Input Voltage Range:90-264VAC

Frequency Range:47-63Hz

Input Current:1.5A at 100VAC

Inrush Current:140A max at 240VAC cold start

Touch Leakage Current(max):≤ 100μA at 264VAC


Environmental

Operating Temperature: -0℃ to 40℃

Storage Temperature:- 20℃ to 60℃

Operating Humidity:10 to 90% RH,non-condensing

Storage Humidity::5 to 95% RH

Operating Altitude:5,000 M


General

Dimensions:76.0(L)×76.0(W)×34.5(H)mm

Weight:260g

MTBF:>50000hrs MIL-HKBK-217 at 25℃

Safety Approvals








Safety Agency/Mark
Medical


Household


CB
IEC60601-1 IEC61558-1

PSE
-J61558-1


TUV-SUD-NRTL
ANSI / AAMI ES60601-1

CAN / CSA-C22.2 NO.60601-1
-

TUV-SUD-MARK


EN60601-1-




 

Model Number

Voltage

(V)

Current

(A)



Rated Power

(W)


Ripple & Noise
(max)
mV



Voltage Tolerance


Line & Load Regulation
EfficiencyStart Up delay
DPS65B1-XXXYYYSPA
11.0-12.00.01-5.0060.00
200
±5%

88.0%
≤3s
12.1-13.00.01-4.6160.00
200±5%

Line: ±1%

Load: ±5%

88.0%
≤3s
13.1-14.00.01-4.2860.00
200±5%
88.0%
≤3s
14.1-15.00.01-4.0060.00
200
±5%
88.0%
≤3s
15.1-16.00.01-3.7560.00
200
±5%
88.0%
≤3s
16.1-17.00.01-3.5260.00
200
±5%
88.0%
≤3s
17.1-18.0

0.01-3.33

60.00
200
±5%
88.0%
≤3s
19.00.01-3.4265.00240
±5%
88.0%
≤3s
19.1-20.00.01-3.2565.00
240±5%
88.0%
≤3s
20.1-21.00.01-3.0965.00
240
±5%
88.0%
≤3s
21.1-22.00.01-2.9565.00
240
±5%
88.0%
≤3s
22.1-23.00.01-2.8265.00
240
±5%
88.0%
≤3s
23.1-24.00.01-2.7165.00
240
±5%
88.0%
≤3s








© Guangdong Degpag Medical Power Supply Co.,Ltd. All Rights Reserved.粤ICP备2021092402号