DPS150-XXXYYYSPA-OP
Date:2023-01-29    Hits:396
  • Power:150W
  • Output voltage:12.00-48.00V
  • Output current:0.01-12.00A
Specification Download

Product Features    

Medical safety approvals    

2 MOPP input to output isolation    

Energy efficiency Level VI    

Touch current ≤ 100μA    

≤0.21W standby power    

Peak load:200% rated load @ 230Vac    

Up to 5000m operating altitude    

Adjustable resistance to adjust output voltage    

Windess design featrues,natural air cooling    


Input

Input Voltage Range:90-264VAC(90-115VAC refer to derating curve)

Frequency Range:47-63Hz

Input Current:2.0A max at 100VAC

Inrush Current:120A max at 240VAC cold start

Touch Leakage Current(max):≤ 100μA at 264VAC


Environmental

Operating Temperature: -30℃ to 70℃(45℃ to 70℃ refer to derating curve)

Storage Temperature:- 40℃ to 80℃

Operating Humidity:5 to 90% RH,non-condensing

Storage Humidity::5 to 95% RH

Operating Altitude:5,000 M

 

General

Dimensions:101.6(L)×50.8(W)×32.0(H)mm

Weight:180g

MTBF:>200000hrs MIL-HKBK-217 at 25℃

Safety Approvals









Safety Agency/Mark
Medical



I.T.E.



CB
IEC60601-1
-

TUV-SUD Mark
EN60601-1
-

NRTL

ANSI/AAMI ES60601-1


-


Model Number

Voltage

(V)

Current

(A)

Rated Power

(W)

Ripple & Noise
(max)
mVpk-pk
Voltage Tolerance
Line & Load Regulation
Efficiency
Start Up delay
DPS150-XXXYYYSPA-OP
12.000.01-11.00132100±5%

92%≤3s
15.000.01-9.00132100Line: ±1%
92%
≤3s
18.000.01-8.00144
120
93%
≤3s
24.000.01-6.00144
120Load: ±5%
93%
≤3s
36.000.01-4.00144
120
94%
≤3s
48.000.01-3.12150
120

94%
≤3s