Language
DP15-150100SPA-OP
Date:2023-01-15    Hits:282
  • Power:15W
  • Output voltage:15V
  • Output current:0.01-1.00A
Specification Download

Product Features    

Medical & ITE safety meet    

2 MOPP input to output isolation    

Efficiency greater than 82%@

Full load and half load    

Touch current ≤ 100μA    

≤0.1W standby power    

Up to 5000m operating altitude    

Input  

Input Voltage Range:90-264VAC

Frequency Range:47-63Hz

Input Current:0.5A max at 100VAC

Inrush Current:100A max at 240VAC cold start

Touch Leakage Current(max):≤ 100μA at 264VAC


Environmental

Operating Temperature: 0℃ to 70℃

Storage Temperature:- 20℃ to 70℃

Operating Humidity:10 to 90% RH,non-condensing

Storage Humidity::5 to 90% RH

Operating Altitude:5,000 M

General

Dimensions:61(L)×38(W)×23.5(H)mm

Weight:40g

MTBF:>100000hrs MIL-HKBK-217 at 25℃


Safety Approvals









Safety Agency/Mark
Medical



I.T.E.(Meet)



CB
IEC60601-1
IEC62368-1



Model Number


Voltage

(V)



Current

(A)



Rated Power

(W)


Ripple & Noise
(max)
mVpk-pk
Voltage Tolerance
Line & Load Regulation
Efficiency
Start Up delay
DP15-XXXYYYSPA-OP
15.00.01-1.0015W200±5%

Line: ±1%

Load: ±5%

82%
≤3s