Language
DP63-XXXYYYSPA-OP
Date:2023-01-15    Hits:355
  • Power:63W
  • Output voltage:18.0V
  • Output current:0.01-3.50A
Specification Download
Product Features
Medical & ITE safety meet
2 MOPP input to output isolation

Energy efficiency Level V


Touch current ≤ 100μA


≤0.3W standby power


Up to 5000m operating altitude

The peak load can reach 7A and last for more than1s



Input

Input Voltage Range:90-264VAC

Frequency Range:47-63Hz

Input Current:1.5A max at 90VAC

Inrush Current:100A max at 240VAC cold start

Touch Leakage Current(max):≤ 100μA at 264VAC


Environmental

Operating Temperature: 0℃ to 40℃

Storage Temperature:- 20℃ to 70℃

Operating Humidity:10 to 90% RH,non-condensing

Storage Humidity::5 to 90% RH

Operating Altitude:5,000 M

 

General

Dimensions:134(L)×64(W)×32(H)mm

Weight:125g

MTBF:>100000hrs MIL-HKBK-217 at 25℃



Safety Approvals









Safety Agency/Mark
Medical



I.T.E.(Meet)



CB
IEC60601-1
IEC62368-1




Model Number


Voltage

(V)



Current

(A)



Rated Power

(W)


Ripple & Noise
(max)
mVpk-pk
Voltage Tolerance
Line & Load Regulation
Efficiency
Start Up delay
DP63-XXXYYYSPA-OP
18.00.01-3.5063W240±5%

Line: ±1%

Load: ±5%

87%
≤3s


© Guangdong Degpag Medical Power Supply Co.,Ltd. All Rights Reserved.粤ICP备2021092402号